Access critical security patches and updates for our products. Browse the latest patch releases, installation instructions, and best practices to ensure your software remains secure and up to date.
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Impacted Software Versions | Actions Required |
---|---|
iScan Control Software 4.0.0 | Configure UCS account credentials. NOTE: The instructions indicate that iScan Control Software v4.0.0 and above require a software update. Version 4.0.0 does not need a software update and only requires configuration of UCS account credentials. |
iScan Control Software 4.0.5 | Contact Illumina Technical Support to upgrade to v4.2.1. |
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Important Notes
Instructions
Patches
Microsoft Release Period | Date Posted by Illumina | Next Scheduled Release by Illumina |
---|---|---|
Q1 2025 |
Q2 2025 | Q3 2025 |
Install in the following order:
1 |
2025-01 Servicing Stack Update for Windows 10 Version 1607 for x64-based Systems (KB5050109) | ||
2 |
2025-01 Cumulative Update for Windows 10 Version 1607 for x64-based Systems (KB5049993) | ||
3 |
2021-01 Update for Windows 10 Version 1607 for x64-based Systems (KB4589210) |
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
iSeq | All | Configure UCS Account Credentials |
This is the software correction for the Local Run Manager security vulnerability of which customers were informed and a temporary security patch was provided in May 2022.
The following table lists the impacted instruments and the required actions for each instrument. Note that all prior versions of control software which utilize Local Run Manager will no longer be available for distribution by Illumina once the software versions listed below are released.
Upon completion of the successful software upgrade, follow Illumina Cybersecurity best practices, including confirming that the firewall is enabled, and Universal Copy Service is running as a standard user.
PQN2024-1496 iSeq and MiniSeq Local Run Manager Cybersecurity Vulnerability
Impacted Instruments and Actions to Take:
Instrument | Action | Release Date |
---|---|---|
iSeq | User to update to iSeq Control Software v3.1 and available Local Run Manager modules (Refer to Additional Information below). | March 13, 2024 |
The new version of iSeq100 Control Software is user-installable; Illumina field service is not required. Install the iSeq100 Control Software v3.1 on your iSeq100 by browsing to the Illumina support site. The page contains the software installer and the document containing software release notes (pages 1-3) and upgrade installation instructions (pages 4-12).
NOTE: MyIllumina log-in is required. For MyIllumina log-in support, contact Illumina Customer Care.
For instruments under an Advanced Exchange Support Plan, onsite field service is not available.
Local Run Manager Modules Compatibility and Availability
iSeq Module | Compatible Version with iSeq Control Software v3.1 | Availability |
---|---|---|
16s Metagenomics |
v2.1.0.33 | Modules available on the Illumina Support Site. |
Amplicon DS | v3.0.0.56 | |
Assembly | v2.1.0.45 | |
DNA Amplicon | v3.0.1.1 | |
DNA Enrichment | v3.1.0.1090 | |
GenerateFASTQ | v3.1.1.56 | |
Library QC | v2.2.0.17 | |
PCR Amplicon | v2.1.0.30 | |
Resequencing | v3.1.0.27 | |
RNA Amplicon | v3.0.1.37 | |
Small RNA | v3.1.0.2505 |
Refer to the following editable PDFs of the recall verification forms. Return the completed form to techsupport@illumina.com to acknowledge receipt of the notification.
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
MiniSeq | MiniSeq Control Software 2.0 andnewer | Configure UCS Account Credentials |
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Secure Communications - Cybersecurity Guidance
This is the software correction for the Local Run Manager security vulnerability of which customers were informed and a temporary security patch was provided in May 2022.
The following table lists the impacted instruments and the required actions for each instrument. Note that all prior versions of control software which utilize Local Run Manager will no longer be available for distribution by Illumina once the software versions listed below are released.
Upon completion of the successful software upgrade, follow Illumina Cybersecurity best practices, including confirming that the firewall is enabled, and Universal Copy Service is running as a standard user.
PQN2024-1496 iSeq and MiniSeq Local Run Manager Cybersecurity Vulnerability
Impacted Instruments and Actions to Take:
Instrument | Action | Release Date |
---|---|---|
MiniSeq | User to update to Miniseq Control Software v2.3 and available Local Run Manager modules (Refer to Additional Information below). | March 13, 2024 |
The new version of Miniseq Control Software is user-installable; Illumina field service is not required. Install the Miniseq Control Software v2.3 on your Miniseq by browsing to the Illumina support site. The page contains the software installer and the document containing software release notes (pages 1-4) and upgrade installation instructions (pages 5-15).
MyIllumina log-in is required. For MyIllumina log-in support, contact Illumina Customer Care.
Local Run Manager Modules Compatibility and Availability
MiniSeq Module | Compatible Version with MiniSeq Control Software v2.3 | Availability |
---|---|---|
16s Metagenomics |
v2.1.0.33 | Modules available on the Illumina Support Site. |
Amplicon DS | v3.0.0.56 | |
Assembly | v2.1.0.45 | |
DNA Amplicon | v3.0.1.1 | |
DNA Enrichment | v3.1.0.1090 | |
GenerateFASTQ | v3.1.1.56 | |
Library QC | v2.2.0.17 | |
PCR Amplicon | v2.1.0.30 | |
Resequencing | v3.1.0.27 | |
RNA Amplicon | v3.0.1.37 | |
RNA Fusion | v3.0.0.25 | |
Small RNA | v3.1.0.2505 | |
TruSight Tumor (TST15) | TBD | Planned for Q2 2024 |
Refer to the following editable PDFs of the recall verification forms. Return the completed form to techsupport@illumina.com to acknowledge receipt of the notification.
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
MiSeq | MiSeq Control Software 4.0 and newer | Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
This is the software correction for the Local Run Manager security vulnerability of which customers were informed and a temporary security patch was provided in May 2022.
The following table lists the impacted instruments and the required actions for each instrument. Note that all prior versions of control software which utilize Local Run Manager will no longer be available for distribution by Illumina once the software versions listed below are released.
Upon completion of the successful software upgrade, follow Illumina Cybersecurity best practices, including confirming that the firewall is enabled, and Universal Copy Service is running as a standard user.
Impacted Instruments and Actions to Take:
Instrument | Action | Release Date |
---|---|---|
MiSeq | User to update to MiSeq Control Software v4.1 and available Local Run Manager modules. (Refer to Additional Informationbelow.) | November 28, 2023 |
The new version of MiSeq Control Software is user-installable, Illumina Field Service is not required. Install the MiSeq Control Software v4.1 on your MiSeq by browsing to the Illumina support site and following the installation instructions provided. Release notes can also be found on the aforementioned Illumina support site. If your MiSeq instrument is on Windows 7, an Illumina Field Service visit is required to upgrade to Windows 10 and the new MiSeq Control Software.
NOTE: MyIllumina log-in is required. For MyIllumina log-in support, contact Illumina Customer Care.
Local Run Manager Modules Compatibility and Availability
MiSeq Module | Compatible Version with MiSeq Control Software v4.1 | Availability |
---|---|---|
16s Metagenomics |
v2.1.0.33 | Modules available on the Illumina Support Site. |
Amplicon DS | v3.0.0.56 | |
Assembly | v2.1.0.45 | |
DNA Amplicon | v3.0.1.1 | |
DNA Enrichment | v3.1.0.1090 | |
GenerateFASTQ | v3.1.1.56 | |
Library QC | v2.2.0.17 | |
PCR Amplicon | v2.1.0.30 | |
Resequencing | v3.1.0.27 | |
RNA Amplicon | v3.0.1.37 | |
RNA Fusion | v3.0.0.25 | |
Small RNA | v3.1.0.2505 | |
TruSight Tumor (TST15) | TBD | Planned for Q2 2024 |
Refer to the following editable PDFs of the recall verification forms. Return the completed form to techsupport@illumina.com to acknowledge receipt of the notification and completion of software update.
Important Notes
Instructions
Patches
Microsoft Release Period | Date Posted by Illumina | Next Scheduled Release by Illumina |
---|---|---|
Q1 2025 |
Q2 2025 | Q2 2025 |
Install in the following order:
1 |
2021-08 Servicing Stack Update for Windows 10 Version 1809 for x64-based Systems (KB5005112) | ||
2 |
2025-01 Cumulative Update for Windows 10 Version 1809 for x64-based Systems (KB5050008) | ||
3 |
2025-01 Cumulative Update for .NET Framework 3.5 and 4.7.2 for Windows 10 Version 1809 for x64 (KB5049608) |
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
MiSeqDx | MiSeq Control Software 4.0 and newer (RUO Mode) | Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
MiSeqDx Operating Software 4.0.1 and newer | Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
This is the software correction for the Local Run Manager security vulnerability of which customers were informed and a temporary security patch was provided in May 2022.
The following table lists the impacted instruments and the required actions for each instrument. Note that all prior versions of control software which utilize Local Run Manager will no longer be available for distribution by Illumina once the software versions listed below are released.
Upon completion of the successful software upgrade, follow Illumina Cybersecurity best practices, including confirming that the firewall is enabled, and Universal Copy Service is running as a standard user.
Impacted Instruments and Actions to Take:
Instrument | Action | Release Date |
---|---|---|
MiSeqDx | Illumina Field Support to update to MiSeqDx Operating Software v4.1 and available Local Run Manager modules. (Refer to Additional Information below.) | November 28, 2023 |
Software installation requires onsite Illumina support. An Illumina Field Representative will contact you to schedule an onsite visit to install the MiSeqDx Operating Software v4.1 (release notes) and the MiSeq Control Software v4.1 (release notes).
Local Run Manager Modules Compatibility and Availability
MiSeqDx Module | Compatible Version with MiSeqDx Control Software v4.1 | Availability |
---|---|---|
CF 139-Variant 2.0 |
v1.0.1.124 | Installed by Illumina Field Representative. |
CF 139-Variant 2.0 Micro | v1.0.0.113 | |
CF Clinical Seq 2.0 | v1.0.1.96 | |
CF Clinical Seq 2.0 Micro | v1.0.0.70 | |
DNA GenerateFASTQ Dx | v1.0.0.299 | |
Germline Variant | v2.0.0.36 | |
Somatic Variant | v2.0.0.12 |
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
NextSeq 1000/2000 | NextSeq Control Software 1.4.1 | Update the system software with NextSeq 1000/2000 Control Software Suite v1.5.0 Installer. |
Important Notes
Instructions
Patches
Microsoft Release Period | Date Posted by Illumina | Next Scheduled Release by Illumina |
---|---|---|
Q1 2025 |
Q2 2025 | Q2 2025 |
Install in the following order:
1 |
2021-08 Servicing Stack Update for Windows 10 Version 1809 for x64-based Systems (KB5005112) | ||
2 |
2025-01 Cumulative Update for Windows 10 Version 1809 for x64-based Systems (KB5050008) | ||
3 |
2025-01 Cumulative Update for .NET Framework 3.5 and 4.8 for Windows 10 Version 1809 for x64 (KB5049615) |
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
NextSeq 500/550 | NextSeq Control Software 4.0 | Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
NextSeq Control Software 4.1 |
Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
This is the software correction for the Local Run Manager security vulnerability of which customers were informed and a temporary security patch was provided in May 2022.
The following table lists the impacted instruments and the required actions for each instrument. Note that all prior versions of control software which utilizeLocal Run Manager will no longer be available for distribution by Illumina once the software versions listed below are released.
Upon completion of the successful software upgrade, follow Illumina Cybersecurity best practices, including confirming that the firewall is enabled, and Universal Copy Service is running as a standard user.
Impacted Instruments and Actions to Take:
Instrument | Action | Release Date |
---|---|---|
NextSeq 500/550 | Illumina Field Support to update NextSeq Control Software v4.2 and available Local Run Manager modules. (Refer to Additional Information below.) | May 31, 2023 |
Software Installation requires onsite Illumina support. An Illumina Field Representative will contact you to schedule an onsite visit to install NextSeq Control Software v4.2.0 (release notes).
Local Run Manager Modules Compatibility Availability
NextSeq 550 Module Description | Compatible Version with NextSeq Control Software v4.2 | Availability |
---|---|---|
16s Metagenomics |
v2.1.0.33 | Modules available on the Illumina Support Site. |
Amplicon DS | v3.0.0.56 | |
Assembly | v2.1.0.45 | |
DNA Amplicon | v3.0.1.1 | |
DNA Enrichment | v3.1.0.1090 | |
GenerateFASTQ | v3.1.1.56 | |
Library QC | v2.2.0.17 | |
PCR Amplicon | v2.1.0.30 | |
Resequencing | v3.1.0.27 | |
RNA Amplicon | v3.0.1.37 | |
RNA Fusion | v3.0.0.25 | |
Small RNA | v3.1.0.2505 | |
TruSight Oncology 500 | v2.3 | Planned for Q3 2023 |
VeriSeq NIPT | v1.0.4.3 | Planned for Q4 2023 |
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
This is the software correction for the Local Run Manager security vulnerability of which customers were informed and a temporary security patch was provided in May 2022.
The following table lists the impacted instruments and the required actions for each instrument. Note that all prior versions of control software which utilize Local Run Manager will no longer be available for distribution by Illumina once the software versions listed below are released.
Upon completion of the successful software upgrade, follow Illumina Cybersecurity best practices, including confirming that the firewall is enabled, and Universal Copy Service is running as a standard user.
Impacted Instruments and Actions to Take:
Instrument | Action | Release Date |
---|---|---|
NextSeq 550Dx | Illumina Field Support to update NextSeqDx Operating Software v1.7 and available Local Run Manager modules (Refer to Additional Information below.) | October 25, 2023 |
Local Run Manager Modules Compatibility and Availability
NextSeq 550Dx Module | Compatible Version with NextSeq 550Dx Operating Software v1.7 | Availability |
---|---|---|
DNA GenerateFASTQ Dx | v1.0.0.299 | Modules available on the Illumina Support Site. |
Germline Variant | v1.3.0.0 | |
Somatic Variant | v1.3.0.0 | |
VeriSeq NIPT | v1.1.2.601 | |
TruSight Oncology Comp IVD | v2.3.7 (IVDD) | Planned for Q4 2023 |
Refer to the following editable PDFs of the recall verification forms. Return the completed form to techsupport@illumina.com to acknowledge receipt of the notification.
Form | Description |
---|---|
FSN2023-1450 NextSeqDx Local Run Manager Update - Verification | Verification Form: Standard Form |
Product Defect Correction Notice (RC-2023-RN-00842-1) | Verification Form: Australia |
PQN2023-1451 NextSeqDx Local Run Manager Update for Japan and Korea-Verification | Verification Form: Japan and Korea |
FSN2023-1450 NextSeqDx Local Run Manager Update - NZ -Verification | Verification Form: New Zealand |
FSN2023-1450 NextSeqDx Local Run Manager Update for Portugal – Verification | Verification Form: Portugal |
FSN2023-1450 NextSeqDx Local Run Manager Update Singapore-Verification | Verification Form: Singapore |
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
NextSeq 550Dx | NextSeq Control Software 4.0 (RUO Mode) | Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
NextSeq Control Software 4.1 |
Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. | |
NextSeq Operating Software 1.0.0-1.3.1 |
Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. | |
NextSeq Operating Software 1.3.3 and newer |
Update software per the actions for this instrument listed in the Local Run Manager Vulnerability Correction patch. |
Important Notes
Instructions
Patches
Microsoft Release Period | Date Posted by Illumina | Next Scheduled Release by Illumina |
---|---|---|
Q4 2024 |
Q4 2024 | Q2 2025 |
Install in the following order:
1 |
2021-08 Servicing Stack Update for Windows 10 Version 1809 for x64-based Systems (KB5005112) | ||
2 |
2024-10 Cumulative Update for Windows 10 Version 1809 for x64-based Systems (KB5044277) | ||
3 |
2024-10 Cumulative Update for .NET Framework 3.5 and 4.7.2 for Windows 10 Version 1809 for x64 (KB5044016) |
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
Thank you for visiting the customer support site for the UCS vulnerability.
The following table lists the impacted instruments and software versions. All instruments require actions outlined in the instruction guide. Illumina recommends that you read the instructions before downloading any software.
NOTE: For any software versions not specified in the table, your instrument is not impacted by this vulnerability and no actions are required. Please return the verification form attached to your notification confirming the action was not performed.
Instrument | Impacted Software Versions | Actions Required |
---|---|---|
NovaSeq 6000 | NovaSeq Control Software 1.7 and older | Configure UCS account credentials. NOTE: The instructions state that the firewall port must be closed for these software versions. Closing the firewall port is not required. |
NovaSeq Control Software 1.8 |
|
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
Thank you for visiting the customer support site for setting a BIOS password on your Illumina system (eg, instrument, server, computer) to keep it secure. Assigning a password to a system’s BIOS provides an added layer of security and is recommended by Illumina, per the Setting BIOS Password CCN2025-1703 sent on March 25, 2025.
Illumina highly recommends that customers assign an Administrator BIOS password on all Illumina systems, regardless of whether they are connected to an external internet or local network.
Refer to the following instructions to create and set an Administrator BIOS password to the following systems:
DRAGEN v3/v4 Server, iScan, iScan Dx, iSeq100, MiniSeq, MiSeq, MiSeqDx, NextSeq 1000/2000, NextSeq 500/550, NextSeq 550Dx, NovaSeq 6000, NovaSeq 6000Dx, NovaSeq X, VeriSeq NIPT Server, as well as any other computers associated with Illumina end-to-end workflow products.
MiSeq i100 comes with an Administrator user password set on the BIOS. As a result, the MiSeq i100 does not require additional steps be taken at this time.
Setting a BIOS Password on Illumina Products (doc # 200068356)
Important Notes
Instructions
Patches
Microsoft Release Period | Date Posted by Illumina | Next Scheduled Release by Illumina |
---|---|---|
Q1 2025 | Q2 2025 | Q2 2025 |
Install in the following order:
1 |
2025-01 Cumulative Update for Windows 10 Version 1809 for x64-based Systems (KB5050008) | ||
2 |
2025-01 Cumulative Update for .NET Framework 3.5 and 4.7.2 for Windows 10 Version 1809 for x64 (KB5049608) |
This Cybersecurity Guidance document contains security requirements which Illumina recommends following to avoid elevation of cybersecurity risks associated with your Illumina product.
Topics covered include:
A MyIllumina account is required to download the listed files. You may be prompted to log in to your MyIllumina account and provide an instrument serial number OR a customer number when selecting a file from the list.
For MyIllumina login support, contact Customer Care.
* For In Vitro Diagnostic Use
Contact an Illumina representative for regional availability.
If you identify a potential vulnerability in an Illumina product, please contact us via email as soon as possible utilizing PGP (Pretty Good Privacy encryption program) as outlined below: